Capabilities & Parameters
Scientific Capabilities
Scientific Application Time-resolved spectroscopy and scattering with ultrafast and ultraintense soft X-rays at the LCLS
Techniques X-ray Emission Spectroscopy (XES)
X-ray Photoelectron Spectroscopy (XPS)
X-ray Absorption Spectroscopy (XAS)
Coherent Imaging
Fourier Transform Holography
Sample Environment A wide range of experimental setups in UHV and UV conditions
Source Parameters
Linac Energy 4.3 GeV for 800 eV 1st harmonic
Undulator 33 Planar hybrid undulator 30 mm gap
Source Size 230 x 230 µm2 FWHM @ 800 eV
Source Divergence 5.7 x 5.7 µrad2 FWHM @ 800 eV
Repetition Rate 120 Hz
Pulse Duration 137 fs
Photons per Pulse ≤1 x 1013
Photon Beam Properties
Sample Configurations Pre-Mono Transmission Sample Chamber, white beam Endstation, white beam Endstation, mono
Focusing Capability (Kirkpatrick-Baez) 1x1 mm (unfocused at sample) < 10x10 um < 10x10 um
Energy Range 500-2000 eV
(initial operations from 825-2000 eV)
grat 1: 500-1500 eV
grat 2: 800-2000 eV
Energy Resolution ~0.1 eV (500 eV) ~0.5 eV (500 eV) ~0.1 eV (500 eV)
~0.2 eV (1000 eV) ~1.0 eV (1000 eV) ~0.2 eV (1000 eV)
~0.6 eV (2000 eV) ~2.0 eV (2000 eV) ~0.6 eV (2000 eV)
Flux (ph/pulse) @ resolution limit 1 x 1013 (500 eV) 1 x 1013 (500 eV) 1.5 x 1011 (500 eV)
8 x 1012 (1000 eV) 8 x 1012 (1000 eV) 2 x 1011 (1000 eV)
4 x 1012 (2000 eV) 4 x 1012 (2000 eV) 4 x 1010 (2000 eV)
Optical Laser Properties
Sample Configurations Pre-Mono Transmission Sample Chamber Endstation
Laser Power 800nm: 7 x 1012 W/cm2 800nm: 2.5 x 1015 W/cm2
400nm: 3 x 1012 W/cm2 400nm: 1 x 1015 W/cm2
Laser Pulse Width 800nm: 35-100 fs 800nm: 35-100 fs
400nm: 50-100 fs 400nm: 50-100 fs
Focusing Capability 1mm (1/e2) 50µm (1/e2)
Diagnostics
Intensity monitor A pulse by pulse intensity monitor after the exit slit for normalizing detector signals. The intensity measurement is relative with a noise level <0.1%
Fast Shutter A fast shutter to quickly turn on or off the beam or to select individual LCLS x-ray pulses. Closes or open in <8 ms. The fast shutter is up stream of the first experimental position.
Alignment Laser A HeNe alignment laser beam coaxial with the FEL beam at the Endstation Position, to facilitate the positioning of samples in the FEL beam. This laser is class IIIa rating or less.